Home/Catalogue/BK-AutoHS96

PCR & extraction

Automatic Nucleic Acid Extractor BK-AutoHS96

BK-AutoHS96

Nucleic-acid preparation, extraction and amplification for molecular-biology workflows.

Made to orderPrice on request
Manufacturer
BIOBASE
Request a quote
Reply within 24 hoursValidated selection and configurationDelivery across mainland Portugal, Madeira, the Azores, and all European countries.

01 / Key features

Key features

  • Molecular-biology workflow
  • Thermal control or automated extraction
  • Configuration matched to required throughput

02 / LaboratoriX technical datasheet

LaboratoriX technical datasheet

Technical parameters consolidated from the manufacturer documentation. Each column corresponds to one model; shared values are clearly identified. Final configuration, accessories, power supply and applicable documentation are confirmed with the quotation.

ParameterBK-AutoHS96
Extraction Method
Magnetic bead method
Working Mode
Automatic sampling+Nucleic acid extraction+PCR reaction system addition
Throughput
1~96, linear slide type sample rack
Extraction Volume
20~1000μl
Processing Time
Complete the processing of 96 samples within 60 minutes(Related to reagents)
Magnetic Bead Recovery
≥98%
Temp Range
RT~105℃, lysis and elution position
Temp Accuracy
0.1℃
Heating Method
Dry bath heating
Heating Speed
RT~100℃≤6min
Shaking Function
Up and down oscillation(1~5 gears adjustable)
Extraction Position
6(96-well deep well plate)
Robotic Arm
A robotic arm for adding samples and reagents
Pipetting Channel
2 channels
Liquid Detection
Pneumatic liquid level detection principle, intelligent detection of blocked needle
Pipetting Tip
50μl, 200μl,1000μl, disposable black conductive needle with filter element
Tip Amount
2~3 tips/sample
Pipetting Accuracy
10μl, CV≤3.0%, Accuracy≤5.0%, 50μl tip
Pipetting Accuracy (2)
50μl, CV≤2.0%, Accuracy≤2.0%, 1000μl tip
Pipetting Accuracy (3)
100μl, CV≤1.5%, Accuracy≤2.0%, 1000μl tip
Sample Volume
2~1000μl
Working Zone
2 pcr positions with cooling function
Working Zone (2)
6 tip positions for three types of tips
Working Zone (3)
2 reagent positions(5ml freezing tube rack position with cooling function, one reserved position)
Protective Function
Start up self-test, Power-off protection, High temperature alarm, Over-temperature protection, Tip removal protection
Disinfection Method
UV lamp(30W*1, 8W*1)
Illumination Lamp
10W LED lamp
Audible Alarm
Yes(Red and blue blinking)
Safety Door Design
With safety lock function, the safety door is opened and the program is suspended
Display
10.1 inch touch screen, Windows system
Scanning
Optional
Interface
LAN interface(Bi-direction LIS optional)
Contamination Control
Built-in air duct and HEPA filter can effectively filter internal aerosols and prevent cross-contamination
IAP Function
Firmware can be upgraded online at any time
Power Supply
110/220V, 50/60Hz
External Size(W*D*H)
1420*850*1842mm
Package Size(W*D*H)
1535*970*1180mm(Main instrument); 1540*970*1160mm(Base)
Gross Weight
360kg(Main instrument); 190kg(Base)

MADE-TO-ORDER SUPPLY

Do you need this equipment?

Share the application, configuration, quantity and delivery location. We prepare a tailored proposal with the required documentation.

Request a quote