Home/Catalogue/BIOBASE2001

ELISA & microplates

Auto ELISA Processor BIOBASE2001

BIOBASE2001

Microplate automation, reading and washing for ELISA assays in research and laboratory diagnostics.

Made to orderPrice on request
Manufacturer
BIOBASE
Request a quote
Reply within 24 hoursValidated selection and configurationDelivery across mainland Portugal, Madeira, the Azores, and all European countries.

01 / Key features

Key features

  • Automated microplate processing
  • Assay-matched reading and washing
  • Configurations for different workloads

02 / LaboratoriX technical datasheet

LaboratoriX technical datasheet

Technical parameters consolidated from the manufacturer documentation. Each column corresponds to one model; shared values are clearly identified. Final configuration, accessories, power supply and applicable documentation are confirmed with the quotation.

ParameterBIOBASE2001
Sample&Reagent Unit — Sample Racks
6*20 push-pull sample rack with detection
Sample&Reagent Unit — Sample Channel
2 channels
Sample&Reagent Unit — Liquid Level Detection
Air pressure level detection, clot detection function
Sample&Reagent Unit — Tip
1000μl, 3 tip racks
Sample&Reagent Unit — Pipetting
10μl~1000μl
Sample&Reagent Unit — Pipetting Precision
100μl: CV≤1%, Accuracy≤2.5% 1000μl: CV≤0.5%, Accuracy≤1%
Sample&Reagent Unit — Reagent Position
4 reagent racks, reagent tanks with detection
Washing Unit — Washer
1 unit
Washing Unit — Washing Head
16-channel(32-pin), double row cleaning, hanging installation, with leakage recovery
Washing Unit — Injection Uniformity
CV<1.5%
Washing Unit — Washing Residual
<1μl
Reading Unit — Reading Channels
8 independent photometric channels
Reading Unit — Linear Range
0.000~3.000Abs
Reading Unit — Repeatability
CV≤1.0%
Reading Unit — Stability
≤±0.003Abs
Reading Unit — Optical Filters
4 standard filters(405, 450, 492, 630nm); Total 10 filters positions
Reading Unit — Reading Accuracy
≤±0.02Abs from 0.0 to 1.0OD ≤±0.03Abs from 1.0 to 2.0OD
Reading Unit — Sensitivity
≥0.01mg/L
Reading Unit — Channel Difference
≤0.02Abs
Incubation Unit — Incubators
8 shaking incubators
Incubation Unit — Cover Positions
1(8 covers)
Incubation Unit — Heating Method
Dry heating with cover
Incubation Unit — Shaking Function
Horizontal circular oscillation, 1~3 levels adjustable, amplitude 2.5mm
Incubation Unit — Temperature Range
RT to 60 °C
Incubation Unit — Temperature Uniformity
0.5℃
Incubation Unit — Temperature Fluctuation
±0.5℃
Data Management — Software
Above Windows 7 system
Data Management — LIS System
Bi-direction, support HL7 protocol
Data Management — Results
Absorbance and results reviewable by software
Data Management — Printer
External optional
Data Management — Built-in Scanner
Optional
Other Function — Biosafety Protection
Optional (Negative pressure working zone, HEPA filter)
Other Function — Disinfection
UV lamp
Other Function — Lighting
LED
Other Function — IAP Function
Firmware can be upgraded online at any time
Working Condition — Power Supply
AC220V±10%, 50/60Hz(Standard); 110V±10%, 60Hz( Optional)
Working Condition — Temperature
10~40℃
Working Condition — Humidity
30~80%
Size&Weight — External Size(W*D*H)
Machine: 1273*750*1650mm Base Stand: 1273*750*777mm
Size&Weight — Net Weight
Machine: 226kg · Base Stand: 98kg
Size&Weight — Package Size(W*D*H)mm
Machine: 1445*920*1130mm Base Stand: 1445*920*1005mm Accessories: 1065*895*510mm
Size&Weight — Gross Weight(kg)
Machine: 300kg Base Stand: 172kg Accessories: 83kg

MADE-TO-ORDER SUPPLY

Do you need this equipment?

Share the application, configuration, quantity and delivery location. We prepare a tailored proposal with the required documentation.

Request a quote